RAMaker Spectrometer
RAMaker
is a professional Raman / PL system which allows multiple excitation
wavelengths on a wide range of laser selection. Maximum of 3 lasers can be
built-in inside the chamber, available wavelengths are: 355, 375, 405, 473,
488, 532, 633, 785 and 830 nm. UV or NIR
sources such as 248, 266 and 325 nm can be mounted externally as well.
It has gained Taiwan invention Patent
(I709732) and USA invention Patent (US 11,340,114 B2), parallel optical design
remains system qualitative without calibration, each laser wavelength is
completely fixed and optimized, the optimized Rayleigh filter is automatically
switched by the software for wavelength correction and measurement.
The switching of different wavelengths is
controlled by the software to achieve the purpose of fully automatic
micro-Raman switching light source without any manual adjustment.
New patent design for visible light
source automatically measuring, software controls the switching
of laser, stable optical system, convenient to do sample co-location
measurement by using different laser lights.
Novel
confocal design including two sets of confocal apertures could precisely focus
to obtain the best signal. The slit is controlled by software and continuously
adjusted from 10um to 2500 μm and 0.8~8 mm. The confocal aperture can
be 10~1000 mm, which can increase the signal resolution by more than 30%.
Continuous neutral density filters
(O.D 2.0~0.04), energy calibration line internally, its error value of energy
control is by less than 1% precisely, build-in and self-input attenuation
ratio; at the same time, the built-in laser can also be set by software at a
rate of 1 mW/
Step to adjust the laser energy output. With the cooperation of the two, the
laser power can be adjusted to 0.01 mW a minimum of output.
Detectors
with high-sensitivity and low-noise can improve system sensitivity. The
third-order peak of silicon wafer can be measured S/N >20:1
RPRM
(Raman Polarization Rotating Mapping) for samples with high crystallinity or
thin films, when it is necessary to measure the difference of polarized Raman,
the software can automatically search for the laser and the angle of the
light-receiving signal; another, the software can automatically find the laser
with the best signal or the setting of zero signal angle, which is convenient
for researchers to observe the lattice arrangement.
The micro-Raman system uses a
5-megapixel CCD and a 5-watt LED light source as the sample image observation
module. The measurement of Raman or observing the surface image of sample is
automatically controlled by the software. When observing the position of the
sample, the laser energy is automatically attenuated less than 0.01%, to
equalize the balance between laser and observed white light brightness; the
standard configuration is 10X, 50X long working distance, 100X, it can be added
5X, 20X, 40X UV objective lens by requirements; when switching each objective
lens, the error of laser light spot on sample position is within 1 μm.
-75 X
50 mm or 100
X 100 mm
-XY
minimum step size 0.01 µm, reproducibility +/-0.02 µm
-Z
minimum step size 0.002 µm
-△
Raman Shift
-With
operating joystick; operating software and video software.
-Containing
automatic regression correction function
-Auto
focusing samples, good image quality