Measurement Ranges:
23°C ± 5°C max. 80%R.H. Accuracy: ±(%rdg +dgt)
Measurement
|
Range
|
Resolution
|
Accuracy (Responding Hz )
|
Input Impedance
|
Max. Input Voltage
|
DE-5003(sine wave)
|
DE-5004*(True RMS)
|
ACV
|
6.000V
|
0.001V
|
0.7%+6(40Hz~1kHz)
|
0.7%+6(40Hz~1kHz)
|
Approx. 11MΩ
|
1000V rms
|
60.00V
|
0.01V
|
0.7%+5(40Hz~1kHz)
|
0.7%+6(40Hz~1kHz)
2.0%+6(1kHz~5kHz)
|
Approx. 10MΩ
|
600.0V
|
0.1V
|
1000V
|
1V
|
1.2%+5(40Hz~1kHz)
|
1.2%+6(40Hz~1kHz)
2.0%+6(1kHz~2kHz)
|
*True RMS: valid from 10%~100% of range
Measurement
|
Range
|
Resolution
|
Accuracy
|
Input Impedance
|
Max. Input Voltage
|
DE-5003
|
DE-5004
|
DCV
|
6.000V
|
0.001V
|
0.1%+3
|
Approx.
11MΩ
|
1000V rms
|
60.00V
|
0.01V
|
0.2%+3
|
Approx. 10MΩ
|
600.0V
|
0.1V
|
0.4%+3
|
1000V
|
1V
|
0.6%+5
|
|
|
|
|
|
|
|
VA+Hz
( at secondary display)
|
Accuracy
|
Max. Input Voltage
|
Range
|
Resolution
|
DE 5003
|
DE 5004
|
600.0Hz
|
0.1Hz
|
0.1%+5
|
1000V rms
|
6.000kHz
|
0.001kHz
|
60.00kHz
|
0.01kHz
|
|
|
|
|
|
Measurement
ADP
|
Range
|
Resolution
|
Accuracy (Responding Hz )
|
Input Impedance
|
Max. Input Voltage
|
DE-5003(sine wave)
|
DE-5004*(True RMS)
|
ACmV
|
600.0mV
|
0.1mV
|
0.7%+5(40Hz~2kHz)
|
0.7%+5(40Hz~10kHz)
|
Approx. 10MΩ
|
1000V rms
|
DCmV
|
600.0mV
|
0.1mV
|
0.2%+5
|
0.2%+5
|
*True RMS: valid from 10%~100% of range
Measurement
|
Range
|
Resolution
|
Accuracy
|
Max. Input Voltage
|
Ω
Resistance
|
600.0 Ω
|
0.1Ω
|
0.5%+5
|
1000V rms
|
6.000kΩ
|
0.001kΩ
|
60.00kΩ
|
0.01kΩ
|
600.0kΩ
|
0.1kΩ
|
6.000MΩ
|
0.001MΩ
|
1.0%+5
|
60.00MΩ
|
0.01MΩ
|
2.5%+5
|
Measurement
|
Range
|
Resolution
|
Remarks
|
Open-loop Voltage
|
Max. Input Voltage
|
Continuity
|
600.0Ω
|
0.1Ω
|
Buzzer emits at around ≦10Ω±3Ω
|
Approx. -3.3V
|
1000V rms
|
Measurement
|
Range
|
Resolution
|
Accuracy
|
Open-loop Voltage
|
Max. Input Voltage
|
Diode
|
6.000V
|
0.001V
|
0.5%+5
|
Approx. 2.8V
|
1000V rms
|
Measurement
|
Range
|
Resolution
|
Accuracy
|
Remarks
|
Max. Input Voltage
|
Hz
|
600.0Hz
|
0.1Hz
|
0.1%+5
|
Available reading of Duty:
5.0%~94.9%(<600Hz)
10.0%~90.0%(<6kHz)
20.0%~80.0%(<60kHz)
Resolution 0.1%
Duty smaller than 5%,
The UL will be show on LCD.
Duty larger than 94.9%,
The OL will be show on LCD.
|
1000V rms
|
6.000kHz
|
0.001kHz
|
60.00kHz
|
0.01kHz
|
600.0kHz
|
0.1kHz
|
6.000MHz
|
0.001 MHz
|
20.00MHz
|
0.01MHz
|
Measurement
|
Range
|
Resolution
|
Accuracy
|
Overload Protection
|
DC μA
|
600.0μA
|
0.1μA
|
0.5%+5
|
630mA/1000V fuse
|
6000μA
|
1μA
|
DC mA
|
60.00mA
|
0.01mA
|
600.0mA
|
0.1mA
|
DC A
|
*10.00A
|
0.01A
|
1.0%+10
|
10A/1000V fuse
|
*overload 11 ~ 20A stands with 30 sec. max.
Measurement
|
Range
|
Resolution
|
Accuracy ( Responding Hz)
|
Overload Protection
|
DE-5003(sine wave)
|
DE-5004**(True RMS)
|
AC μA
|
600.0μA
|
0.1μA
|
1.0%+10(40Hz~1kHz)
1.5%+10(1kHz~2kHz)
|
1.0%+10(40Hz~1kHz)
1.5%+10(1kHz~5kHz)
|
630mA/1000V fuse
|
6000μA
|
1μA
|
AC mA
|
60.00mA
|
0.01mA
|
600.0mA
|
0.1mA
|
AC A*
|
10.00A
|
0.01A
|
1.5%+10(40Hz~2kHz)
|
1.5%+10(40Hz~5kHz)
|
10A/1000V fuse
|
*over load 11 ~ 20A stands with 30 sec. max.
**True RMS: valid from 10%~100% of range
LCR Inductance
Ls / Lp
|
0.3∼60.0μH
|
60.1∼600.0μH
|
601∼6000μH
|
6.01∼60.00mH
|
60.1∼600.0mH
|
601∼6000mH
|
6.01∼60.00H
|
60.1∼100.0H
|
Resolution
|
0.1μH
|
0.1μH
|
1μH
|
0.01mH
|
0.1mH
|
1mH
|
0.01H
|
0.1H
|
Accuracy
|
±(3.0%+20)
|
±(1.5%+5)
|
±(1.0%+5)
|
±(1.0% +5)
|
±(1.0%+5)
|
±(1.0%+5)
|
±(1.5%+5)
|
±(3.0%+10)
|
LCR Capacitance
Cs / Cp
|
3.0∼600.0pF
|
601∼6000pF
|
6.01∼60.00nF
|
60.1∼600.0nF
|
601∼6000nF
|
6.01∼60.00μF
|
60.1∼600.0μF
|
0.60∼6.00mF
|
Resolution
|
0.1pF
|
1pF
|
0.01nF
|
0.1nF
|
1nF
|
0.01μF
|
1μF
|
0.01mF
|
Accuracy
|
±(2.0%+20)
|
±(1.0%+5)
|
±(0.3%+5)
|
±(0.3%+5)
|
±(0.5%+5)
|
±(0.8%+5)
|
±(1.5%+10)
|
±(3.0%+20)
|
LCR Resistance
Rs / Rs
|
0.1∼60.00Ω
|
60.1∼600.0Ω
|
601∼6000Ω
|
6.01∼60.00KΩ
|
60.1∼600.0KΩ
|
601∼6000KΩ
|
6.01∼20.00MΩ
|
Resolution
|
0.01Ω
|
0.1Ω
|
1Ω
|
0.01KΩ
|
0.1KΩ
|
1KΩ
|
0.01MΩ
|
Accuracy
|
±(1.0%+20)
|
±(0.3%+5)
|
±(0.3%+5)
|
±(0.3%+5)
|
±(0.3%+5)
|
±(1.5%+5)
|
±(6.0%+10)
|
*Above accuracy is guaranteed after open/short calibration.
*Some types of inductor, the inductance value changes acutely according to the test voltage applied.
*The meter provides the output voltage around 2.2kΩ / 0.6Vrms at sine wave. However, the test voltage is altered in according to the resistance of the inductor be tested.
*Temperature coefficient:0.1 x (specified accuracy) / ℃ (from 0℃ to 18℃ or 28℃ to 50℃)
TEMP./Humidity:
-
Operation:0°C~ 50°C(32℉~122℉)/<80% R.H.(no condensation)
-
Storage:-10°C~ 60°C(14℉~140℉)/<70%R.H.(no condensation)
Dimension & Weight:
-
181(L)x91(W)x57(H)mm
-
Approx.400g (excluding battery)
Accessories:
-
1.5V AA Battery x 2
-
Instruction Manual x 1
-
Test Lead (red+black) x 1
-
Test Alligator Clip (red+black) x 1